Teledyne API - Model T200H/T200M Operation Manual
Getting Started
45
Source of
SAMPLE GAS
Removed during
calibration
Instrument
Chassis
SAMPLE
EXHAUST
PUMP
MODEL 701
Zero Gas
Generator
V
E
N
T
3-way Valve
Manual
Control Valve
VENT
here if input
is pressurized
NO
X
Gas
(High Concentration)
Figure 3-14:
Pneumatic Connections–Basic Configuration–Using Bottled Span Gas
1. Attach a 1/4" exhaust line between the external pump and the EXHAUST port of the
analyzer.
2. Attach an additional 1/4" exhaust port of the pump.
CAUTION
The exhaust from the analyzer must be vented outside the shelter or immediate area
surrounding the instrument and conform to all safety requirements using a maximum
of 10 meters of 1/4” PTFE tubing.
3. Attach a sample inlet line to the SAMPLE inlet port. Ideally, the pressure of the
sample gas should be equal to ambient atmospheric pressure.
Note
Maximum pressure of any gas at the SAMPLE inlet should not exceed 1.5 in-Hg
above ambient pressure and ideally should equal ambient atmospheric pressure.
In applications where the sample gas is received from a pressurized manifold, a
vent must be provided to equalize the sample gas with ambient atmospheric
pressure before it enters the analyzer.
The vented gas must be routed outside the immediate area or shelter
surrounding the instrument.
4. Once the appropriate pneumatic connections have been made, check all pneumatic
fittings for leaks using procedures defined in Section 7.5.1.
07270D DCN7141
Содержание T200H
Страница 2: ......
Страница 4: ...Teledyne API Model T200H T200M Operation Manual ii This page intentionally left blank 07270D DCN7141 ...
Страница 8: ...Teledyne API Model T200H T200M Operation Manual vi This page intentionally left blank 07270D DCN7141 ...
Страница 10: ...Teledyne API Model T200H T200M Operation Manual viii This page intentionally left blank 07270D DCN7141 ...
Страница 12: ...Teledyne API Model T200H T200M Operation Manual x This page intentionally left blank 07270D DCN7141 ...
Страница 344: ...This page intentionally left blank B 2 07270D DCN7141 ...
Страница 353: ...APPENDIX D 07270D DCN7141 07270D DCN7141 D 1 D 1 ...