Teledyne API - Model T200H/T200M Operation Manual
Principles of Operation
305
8.8.2. CALIBRATION - SLOPE AND OFFSET
Aside from the hardware calibration of the preamplifier board (Section 7.6.4) upon
factory checkout, calibration of the analyzer is usually performed in software. During
instrument calibration (Section 5) the user enters expected values for span gas
concentration through the front panel keypad and supplies the instrument with sample gas
of know NO and NO
X
concentrations. The readings are then compared to the expected
values and the software computes values for the new instrument slope and offset for both
NO and NO
X
response. These values are stored in memory for use in calculating the NO,
NO
X
and NO
2
concentration of the sample gas. By default, the DAS stores 200 software
calibration settings for documentation, review and data analysis.
Instrument slope and offset values recorded during the last calibration can be viewed on
the front panel.
NO SLOPE
,
NOX SLOPE
,
NO OFFS
and
NOX OFFS
are four of the
test parameters accessible through the
<TST TST>
buttons.
8.8.3. TEMPERATURE/PRESSURE COMPENSATION (TPC)
The software features a compensation of some temperature and pressure changes critical
in the measurement of NO and NO
X
concentration. When the TPC feature is enabled
(default setting), the analyzer divides the value of the PMT output signal (
PMTDET)
by
a value called
TP_FACTOR
.
TP_FACTOR
is calculated according to the following
equation.
×
×
×
=
)
(
)
(
)
(
(
)
(
)
(
7
)
(
)
(
_
K
298
K
TEMP
BOX
D
Hg
in
29.92
Hg
in
SAMP
C
Hg
in
RCEL
Hg
in
B
K
323
K
TEMP
RCELL
A
FACTOR
TP
(Equation 9-5)
Where A, B, C, D are gain functions. The four parameters used to compute TP_FACTOR
are:
•
RCELL TEMP
: The temperature of the reaction cell, measured in K.
•
RCEL
: The pressure of the gas in the vacuum manifold, measured in in-Hg-A.
•
SAMP
: The pressure of the sample gas before it reaches the reaction cell, measured
in in-Hg-A. This measurement is ~1 in-Hg-A lower than atmospheric pressure.
•
BOX TEMP
: The temperature inside the analyzer’s case measured in K. This is
typically about 5 K higher than room temperature.
The current value of all four of these measurements are viewable as
TEST FUNCTIONS
through the instrument’s front panel display.
that, as
RCEL TEMP, BOX TEMP
and
SAMP
pressure increase, the value of
TP_FACTOR
increases and, hence, the PMTDET value decreases. Conversely,
increases in the reaction cell pressure (
RCEL
) decrease
TP_FACTOR
and, hence
increase the
PMTDET
value. These adjustments are meant to counter-act changes in the
concentrations caused by these parameters.
07270D DCN7141
Содержание T200H
Страница 2: ......
Страница 4: ...Teledyne API Model T200H T200M Operation Manual ii This page intentionally left blank 07270D DCN7141 ...
Страница 8: ...Teledyne API Model T200H T200M Operation Manual vi This page intentionally left blank 07270D DCN7141 ...
Страница 10: ...Teledyne API Model T200H T200M Operation Manual viii This page intentionally left blank 07270D DCN7141 ...
Страница 12: ...Teledyne API Model T200H T200M Operation Manual x This page intentionally left blank 07270D DCN7141 ...
Страница 344: ...This page intentionally left blank B 2 07270D DCN7141 ...
Страница 353: ...APPENDIX D 07270D DCN7141 07270D DCN7141 D 1 D 1 ...