2. SPECIFICATIONS AND APPROVALS
2.1. T200H/M OPERATING SPECIFICATIONS
Table 2-1: Model T200H/M Basic Unit Specifications
Parameter
Specification
Min/Max Range
(Physical Analog Output)
T200H
:
Min: 0-5 ppm Full scale
Max: 0-5000 ppm Full scale (selectable,
independent NO, NO
2
, NO
X
ranges and
auto ranging supported)
T200M
:
Min: 0-1 ppm Full scale
Max: 0-200 ppm Full scale (selectable,
independent NO, NO
2
, NO
X
ranges and
auto ranging supported)
Measurement Units
ppm, mg/m
3
(user selectable)
Zero Noise
<20 ppb (RMS)
Span Noise
<0.2% of reading (RMS) above 20 ppm
Lower Detectable Limit
40 ppb (2x noise as per USEPA)
Zero Drift (24 hours)
<20 ppb (at constant temperature and voltage.)
Span Drift (24 hours)
<0.5% of reading (at constant temperature and voltage.)
Lag Time
20 s
Rise/Fall Time
95% in <60 s (in switching mode)
Linearity
1% of full scale
Precision
0.5% of reading above 5ppm
Gas Flow Rates
T200H
:
•
40 cm³/min sample gas through NO
2
converter & sensor module
•
250 cm
3
/min ± 10% through bypass
manifold
•
290 cm³/min total flow ± 10%
T200M
:
250 cm³/min sample gas through NO
2
converter & sensor module ± 10%
O
2
Sensor option adds 80 cm³/min to total flow through T200H/M when installed.
AC Power
T200H
:
Rating
Typical Power Consumption
110V-120V, 60 Hz 3.0A
220V-240V, 50/60 Hz 3.0A
175W
155W
T200M
:
110V-120V, 60 Hz 3.0A
220V-240V, 50/60 Hz 3.0A
55W
75W
Recorder Offset
± 10%
Environmental
Installation category (over-voltage category) II; Pollution degree 2
Intended for Indoor Use Only at Altitudes
≤
2000m
Analog Outputs
4 user configurable outputs
Analog Output Ranges
All Outputs: 0.1 V, 1 V, 5 V or 10 V
Three outputs convertible to 4-20 mA isolated current loop.
All Ranges with 5% under/over-range
Analog Output Resolution
1 part in 4096 of selected full-scale voltage (12 bit)
Status Outputs
8 Status outputs from opto-isolators, 7 defined, 1 spare
Control Inputs
6 Control inputs, 4 defined, 2 spare
Alarm outputs
2 relay alarms outputs (Optional equipment) with user settable alarm limits
- 1 Form C: SPDT; 3 Amp @ 125 VAC
Temperature Range
5 - 40
°
C operating range
Humidity Range
0-95% RH non-condensing
07270D DCN7141
Содержание T200H
Страница 2: ......
Страница 4: ...Teledyne API Model T200H T200M Operation Manual ii This page intentionally left blank 07270D DCN7141 ...
Страница 8: ...Teledyne API Model T200H T200M Operation Manual vi This page intentionally left blank 07270D DCN7141 ...
Страница 10: ...Teledyne API Model T200H T200M Operation Manual viii This page intentionally left blank 07270D DCN7141 ...
Страница 12: ...Teledyne API Model T200H T200M Operation Manual x This page intentionally left blank 07270D DCN7141 ...
Страница 344: ...This page intentionally left blank B 2 07270D DCN7141 ...
Страница 353: ...APPENDIX D 07270D DCN7141 07270D DCN7141 D 1 D 1 ...