Troubleshooting & Service
Teledyne API - Model T200H/T200M Operation Manual
240
7.5.4.1. AC Configuration – Internal Pump (JP7)
AC power configuration for internal pumps is set using Jumper set JP7 (see Figure 7-4
for the location of JP7).
Table 7-3: AC Power Configuration for Internal Pumps (JP7)
LINE
POWER
LINE
FREQUENCY
JUMPER
COLOR
FUNCTION
JUMPER
BETWEEN
PINS
110VAC
115 VAC
60 HZ
WHITE
Connects pump pin 3 to 110 / 115 VAC power line
2 to 7
Connects pump pin 3 to 110 / 115 VAC power line
3 to 8
Connects pump pins 2 & 4 to Neutral
4 to 9
50 HZ
1
BLACK
Connects pump pin 3 to 110 / 115 VAC power line
2 to 7
Connects pump pin 3 to 110 / 115 VAC power line
3 to 8
Connects pump pins 2 & 4 to Neutral
4 to 9
220VAC
240 VAC
60 HZ
BROWN
Connects pump pins 3 and 4 together
1 to 6
Connects pump pin 1 to 220 / 240VAC power line
3 to 8
50 HZ
1
BLUE
Connects pump pins 3 and 4 together
1 to 6
Connects pump pin 1 to 220 / 240VAC power line
3 to 8
1
A jumper between pins 5 and 10 may be present on the jumper plug assembly, but has no function on the T200H/M
analyzers.
Figure 7-12:
Pump AC Power Jumpers (JP7)
07270D DCN7141
Содержание T200H
Страница 2: ......
Страница 4: ...Teledyne API Model T200H T200M Operation Manual ii This page intentionally left blank 07270D DCN7141 ...
Страница 8: ...Teledyne API Model T200H T200M Operation Manual vi This page intentionally left blank 07270D DCN7141 ...
Страница 10: ...Teledyne API Model T200H T200M Operation Manual viii This page intentionally left blank 07270D DCN7141 ...
Страница 12: ...Teledyne API Model T200H T200M Operation Manual x This page intentionally left blank 07270D DCN7141 ...
Страница 344: ...This page intentionally left blank B 2 07270D DCN7141 ...
Страница 353: ...APPENDIX D 07270D DCN7141 07270D DCN7141 D 1 D 1 ...