Troubleshooting & Service
217
To view or clear a warning messages press:
SAMPLE
SYSTEM RESET
NOX = XXX.X
TEST
CAL
MSG CLR SETUP
If warning messages reappear,
perform a combined error analysis
until the problem is resolved. Do
not repeatedly clear warnings
without corrective action.
Press
CLR
to clear the current
warning message. If more than
one warning is active, the next
message will take its place
.
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL
MSG
CLR SETUP
SAMPLE
SYSTEM RESET
NOX = XXX.X
< TST TST > CAL
MSG
CLR
SETUP
<TST TST
> buttonss replaced
with
TEST
button. Pressing
TEST
suppresss warning messages.
MSG
indicates that warning
messages are active.
Figure 7-1:
Viewing and Clearing Warning Messages
7.1.2. FAULT DIAGNOSIS WITH TEST FUNCTIONS
Besides being useful as predictive diagnostic tools, the TEST functions, viewable from
the front panel, can be used to isolate and identify many operational problems when
combined with a thorough understanding of the analyzer’s theory of operation (Section
8). We recommend using the APICOM remote control program to download, graph and
archive TEST data for analysis and long-term monitoring of diagnostic data (Section
4.15.2.8).
The acceptable ranges for these test functions are listed in Appendix A-3. The actual
values for these test functions on checkout at the factory were also listed in the
Final
Test and Validation Data Sheet
, which was shipped with the instrument. Values outside
the acceptable ranges indicate a failure of one or more of the analyzer’s subsystems.
Functions with values that are within the acceptable range but have significantly
changed from the measurements recorded on the factory data sheet may also indicate a
failure or a maintenance item. A problem report worksheet has been provided in
Appendix C (Teledyne API part number 04503) to assist in recording the value of these
test functions. The following table contains some of the more common causes for these
values to be out of range.
07270D DCN7141
Содержание T200H
Страница 2: ......
Страница 4: ...Teledyne API Model T200H T200M Operation Manual ii This page intentionally left blank 07270D DCN7141 ...
Страница 8: ...Teledyne API Model T200H T200M Operation Manual vi This page intentionally left blank 07270D DCN7141 ...
Страница 10: ...Teledyne API Model T200H T200M Operation Manual viii This page intentionally left blank 07270D DCN7141 ...
Страница 12: ...Teledyne API Model T200H T200M Operation Manual x This page intentionally left blank 07270D DCN7141 ...
Страница 344: ...This page intentionally left blank B 2 07270D DCN7141 ...
Страница 353: ...APPENDIX D 07270D DCN7141 07270D DCN7141 D 1 D 1 ...