Setup Mode Menus
Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
86
5.4.3.2. SETUP
RNGE
MODE
IND: CONFIGURING THE T200 ANALYZER FOR
INDEPENDENT RANGE MODE
The independent range mode (
IND
) assigns the three NO
x
, NO and NO
2
concentrations
to individual analog output channels. In
IND
range mode the
RANGE
test function
displayed on the front panel will then be replaced by three separate functions:
Table 5-1:
IND Mode Analog Output Assignments
TEST
FUNCTION
CONCENTRATION
REPORTED
ANALOG OUTPUT
CHANNEL
RANGE1
NO
x
A1
RANGE2
NO
A2
RANGE3
NO
2
A3
Each can be configured with a different reporting range upper limit and analog signal
span:
EXAMPLE:
NO
x
Concentration –
RANGE1
Set for 0-800 ppb & output
A1
set for 0-10 VDC
NO
Concentration –
RANGE2
Set for 0-200 ppb & output
A2
set for 0-5 VDC
NO
2
Concentration –
RANGE3
Set for 0-400 ppb & output
A3
set for 0-5 VDC
Setting analog range limits to different values does not affect the instrument‟s
calibration.
06858E DCN7057
Содержание T200
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Страница 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Страница 344: ...Index Teledyne API T200 NO NO2 NOX Analyzer Operation Manual 342 This page intentionally left blank 06858E DCN7057 ...
Страница 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Страница 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Страница 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Страница 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Страница 393: ...06858E DCN7057 D 7 ...