
Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
Getting Started
57
3.3.2.4. CONNECTIONS W/AMBIENT ZERO/PRESSURIZED SPAN VALVES (OPT 50B)
This calibration valve package is appropriate for applications where Span Gas is being
supplied from a pressurized source such as bottled NIST SRM gases. This option
includes:
a critical flow orifice and vent to maintain the span Gas supply at 1 ATM
a shutoff valve to preserve the span gas source when it is not in use
two solenoid valves for the user to switch either zero, span or sample gas to the
instrument’s sensor
three additional gas inlet ports (
ZERO AIR,
SPAN
and
VENT
)
Figure 3-23:
Rear Panel Layout with Ambient Zero/Pressurized Span Valves OPT 50B
06858E DCN7057
Содержание T200
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Страница 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Страница 344: ...Index Teledyne API T200 NO NO2 NOX Analyzer Operation Manual 342 This page intentionally left blank 06858E DCN7057 ...
Страница 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Страница 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Страница 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Страница 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Страница 393: ...06858E DCN7057 D 7 ...