Troubleshooting & Service
Teledyne API
– T200 NO/NO2/NOx Analyzer Operation Manual
270
Concentration display
shows all gasses.
SAMPLE
RANGE=500.0 PPB NOX= XXXX
<TST
TST>
CAL
SETUP
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
DIAG
SIGNAL I/O
PREV
NEXT
ENTR EXIT
SETUP X.X
ENTER PASSWORD
8
1
8
ENTR
EXIT
Continue pressing
NEXT
until ...
DIAG OZONE GEN OVERRIDE
PREV
NEXT
ENTR
EXIT
Continue pressing
<TST
or
TST>
until ...
SAMPLE
PMT=2750 MV NOX= XXXX
<TST TST> CAL
SETUP
DIAG OZONE
OZONE GEN OVERIDE
OFF
EXIT
Toggling this button
turns
ON/OFF
the O
3
generator.
Note
This is one of the two settings in the DIAG menu that is retained after you exit the
menu.
12.7.16. INTERNAL SPAN GAS GENERATOR AND VALVE OPTIONS
The zero/span valves and internal span gas generator options need to be enabled in the
software (contact the factory on how to do this).
Check for the physical presence of the valves or the IZS option.
Check front panel for correct software configuration. When the instrument is in
SAMPLE mode, the front panel display should show CALS and CALZ buttons in the
second line of the display. The presence of the buttons indicates that the option
has been enabled in software. In addition, the IZS option is enabled if the TEST
functions show a parameter named IZS TEMP.
The semi-permeable PTFE membrane of the permeation tube is severely affected by
humidity. Variations in humidity between day and night are usually enough to yield
very variable output results. If the instrument is installed in an air-conditioned shelter,
the air is usually dry enough to produce good results. If the instrument is installed in an
environment with variable or high humidity, variations in the permeation tube output
06858E DCN7057
Содержание T200
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Страница 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Страница 344: ...Index Teledyne API T200 NO NO2 NOX Analyzer Operation Manual 342 This page intentionally left blank 06858E DCN7057 ...
Страница 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Страница 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Страница 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Страница 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Страница 393: ...06858E DCN7057 D 7 ...