Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
Troubleshooting & Service
255
12.7.7. PRESSURE / FLOW SENSOR ASSEMBLY
The flow and pressure sensors of the T200 are located on a PCA just behind the PMT
sensor (see Figure 3-5) can be checked with a voltmeter.
Figure 12-6:
Pressure / Flow Sensor Assembly
The following procedure assumes that the wiring is intact and that the motherboard and
power supplies are operating properly:
12.7.7.1. BASIC PCA OPERATION CHECK:
Measure the voltage between TP2 and TP1 C1 it should be 10 VDC ± 0.25 VDC. If
not then the board is bad. Replace the PCA.
12.7.7.2. SAMPLE PRESSURE SENSOR CHECK:
1. Measure the pressure on the inlet side of S1 with an external pressure meter.
2. Measure the voltage across TP4 and TP1.
The expected value for this signal should be:
Expected mVDC
4660
mvDC
250
mvDC
+
x
(
)
=
Pressure
30.0
Hg-In-A
± 10%
rdg
EXAMPLE: If the measured pressure is 20 Hg-in-A, the expected voltage level between
TP4 and TP1 would be between 2870 mVDC and 3510 mVDC.
EXAMPLE: If the measured pressure is 25 Hg-in-A, the expected voltage level between
TP4 and TP1 would be between 3533 mVDC and 4318 mVDC.
06858E DCN7057
Содержание T200
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Страница 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Страница 344: ...Index Teledyne API T200 NO NO2 NOX Analyzer Operation Manual 342 This page intentionally left blank 06858E DCN7057 ...
Страница 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Страница 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Страница 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Страница 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Страница 393: ...06858E DCN7057 D 7 ...