11.2. PREDICTIVE DIAGNOSTICS
Predictive diagnostic functions, including failure warnings and alarms built into the
analyzer‟s firmware, aid in determining whether and when repairs are necessary.
The Test Functions can also be used to predict failures by looking at how their values
change over time, compared to the values recorded on the printed record of the
Final
Test and Validation Data Sheet
, P/N 04490. The internal data acquisition system (DAS)
is a convenient way to record and track these changes. Use APICOM (Section 8.1.1) to
download and review this data from a remote location.
The following table, checked weekly, can be used as a basis for taking action as these
values change with time.
Table 11-2: Predictive Uses for Test Functions
FUNCTION
EXPECTED
ACTUAL
INTERPRETATION & ACTION
RCEL
(pressure)
Constant to within
± 0.5 in-Hg-A
Fluctuating
Developing leak in pneumatic system. Check for leaks.
Slowly increasing
Pump performance is degrading. Rebuild pump when pressure
is above 10 in-Hg-A.
SAMP
(pressure)
Constant within
atmospheric
changes
Fluctuating
Developing leak in pneumatic system. Check for leaks.
Slowly increasing
Flow path is clogging up. Replace orifice filters.
Slowly decreasing
Developing leak in pneumatic system to vacuum (developing
valve failure). Check for leaks.
OZONE FL
Constant to within
± 15
Slowly decreasing
Flow path is clogging up. Replace orifice filters.
AZERO
Constant within
±20 of check-out
value
Significantly increasing
Developing AZERO valve failure. Replace valve.
PMT cooler failure. Check cooler, circuit, and power supplies.
Developing light leak.
O
3
air filter cartridge is exhausted. Change chemical.
NO
2
(Concentration)
Constant for
constant
concentrations
Slowly decreasing
signal for same
concentration
Converter efficiency may be degrading. Replace converter
components.
NO
2
with IZS Option
installed
(Concentration)
Constant
response from day
to day
Decreasing over time
Change in instrument response. Low level (hardware) calibrate
the sensor.
Degradation of IZS permeation tube. Change permeation tube.
Heavily fluctuating from
day to day
Ambient changes in moisture are affecting the performance.
Add a dryer to the zero air inlet.
NO
(Concentration)
Constant for
constant
concentration
Decreasing over time
Drift of instrument response; clean RCEL window. Check for
flow leaks or irregularities.
06858E DCN7057
Содержание T200
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Страница 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Страница 344: ...Index Teledyne API T200 NO NO2 NOX Analyzer Operation Manual 342 This page intentionally left blank 06858E DCN7057 ...
Страница 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Страница 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Страница 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Страница 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Страница 393: ...06858E DCN7057 D 7 ...