Calibration Procedures
Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
194
9.4.2. MANUAL CALIBRATION CHECKS WITH VALVE OPTIONS
INSTALLED
SAMPLE
NOX STB= XXX.X PPB NOX= XXXX
<TST TST> CAL CALZ
CALS
SETUP
SAMPLE
NOX STB= XXX.X PPB NOX= XXXX
<TST TST> CAL
CALZ
CALS
SETUP
Set the Display to show
the NOX STB test
function.
This function calculates
the stability of the NO/NO
x
measurement.
Toggle TST> button until ...
Analyzer display
continues to cycle
through all of the
available gas
measurements
throughout this
procedure.
SAMPLE
RANGE=500.0 PPB NOX= XXXX
<TST
TST>
CAL CALZ CALS
SETUP
The ZERO and/or SPAN
buttons will appear at various
points of this process.
It is not necessary to press
them.
Wait until NOX STB
falls below 0.5 PPB.
This may take several
minutes.
Wait until NOX STB
falls below 0.5 PPB.
This may take several
minutes.
Record NO
X
, NO or NO
2
zero point readings
DO NOT PRESS THE ENTR KEY
Record NO
X
, NO, NO
2
span point readings
DO NOT PRESS THE ENTR button
06858E DCN7057
Содержание T200
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Страница 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Страница 344: ...Index Teledyne API T200 NO NO2 NOX Analyzer Operation Manual 342 This page intentionally left blank 06858E DCN7057 ...
Страница 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Страница 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Страница 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Страница 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Страница 393: ...06858E DCN7057 D 7 ...