Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
Calibration Procedures
193
9.4. MANUAL CALIBRATION AND CAL CHECKS WITH THE
VALVE OPTIONS INSTALLED
There are a variety of valve options available on the T200 for handling calibration gases
(see Section 1.3 for descriptions of each).
Generally performing calibration checks and zero/span point calibrations on analyzers
with these options installed is similar to the methods discussed in the previous sections.
The primary differences are:
On instruments with Z/S valve options, zero air and span gas is supplied to the
analyzer through other gas inlets besides the sample gas inlet.
The zero and span calibration operations are initiated directly and independently
with dedicated buttons (
CALZ
&
CALS
).
9.4.1. SETUP FOR CALIBRATION USING VALVE OPTIONS
Each of the various calibration valve options requires a different pneumatic setup that is
dependent on the exact nature and number of valves present. Refer to the following
diagrams for information on each or these valve sets.
06858E DCN7057
Содержание T200
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Страница 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Страница 344: ...Index Teledyne API T200 NO NO2 NOX Analyzer Operation Manual 342 This page intentionally left blank 06858E DCN7057 ...
Страница 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Страница 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Страница 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Страница 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Страница 393: ...06858E DCN7057 D 7 ...