9.2.3.2. ZERO/SPAN POINT CALIBRATION PROCEDURE
Press
ENTR
to change
the
OFFSET
&
SLOPE
values based on the zero
point measurement.
Press
EXIT
to leave the
calibration unchanged and
return to the previous
menu.
SAMPLE
STABIL=XXXX PPB
NOX= XXXX
< TST TST > CAL SETUP
M-P CAL
STABIL=XXXX PPB
NOX= XXXX
<TST TST>
ZERO
CONC
EXIT
EXIT
at this point
returns to the
SAMPLE menu.
Press
ENTR
to change
the
OFFSET
&
SLOPE
values based on the zero
point measurement.
Press
EXIT
to leave the
calibration unchanged and
return to the previous
menu.
Allow zero gas to enter the sample port
at the rear of the analyzer.
Wait until
STABIL
falls below
1.0 PPB.
This may take several
minutes.
SAMPLE
STABIL=XXXX PPB
NOX= XXXX
< TST TST >
CAL
SETUP
M-P CAL
STABIL=XXXX PPB
NOX= XXXX
<TST TST>
ENTR
CONC
EXIT
M-P CAL
STABIL=XXXX PPB
NOX= XXXX
<TST TST> ZERO
SPAN
CONC
EXIT
Allow span gas to enter the sample port
at the rear of the analyzer.
Wait until
STABIL
falls below
1.0 PPB.
This may take several
minutes.
M-P CAL
STABIL=XXXX PPB
NOX= XXXX
< TST TST >
CAL
SETUP
M-P CAL
STABIL=XXXX PPB
NOX= XXXX
<TST TST>
ENTR
CONC
EXIT
M-P CAL
STABIL=XXXX PPB
NOX= XXXX
<TST TST> ENTR
CONC
EXIT
The
SPAN
button now
appears during the transition
from zero to span.
You may see both buttons.
If either the
ZERO
or
SPAN
buttons fail to appear refer to
the Troubleshooting section
of this manual.
SAMPLE RANGE=500.0 PPB NOX= XXXX
<TST
TST>
CAL
SETUP
Set the Display to show
the
STABIL
test function.
This function calculates
the stability of the NOX
measurement.
Toggle
TST>
button until ...
Note
If the ZERO or SPAN buttons are not displayed, the measurement made during is
out of the allowable range allowed for a reliable calibration. See Section 12 for
troubleshooting tips.
06858E DCN7057
Содержание T200
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Страница 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Страница 344: ...Index Teledyne API T200 NO NO2 NOX Analyzer Operation Manual 342 This page intentionally left blank 06858E DCN7057 ...
Страница 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Страница 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Страница 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Страница 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Страница 393: ...06858E DCN7057 D 7 ...