Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485)
A-21
APPENDIX A-3: Warnings and Test Measurements, S/W Version C.7 (E-Series), 1.0.5
(T-Series)
Table A-2: Warning Messages
Name
1
Message Text
Description
Warnings
WSYSRES
SYSTEM RESET
Instrument was power-cycled or the
CPU was reset.
WDATAINIT DATA
INITIALIZED Data storage was erased.
WCONFIGINIT
CONFIG INITIALIZED
Configuration storage was reset to
factory configuration or erased.
WSO2ALARM1
7
SO2 ALARM 1 WARN
SO
2
concentration alarm limit #1
exceeded
WSO2ALARM2
7
SO2 ALARM 2 WARN
SO
2
concentration alarm limit #2
exceeded
WH2SALARM1
7
H2S ALARM 1 WARN
H
2
S/TRS concentration alarm limit
#1 exceeded
WH2SALARM2
7
H2S ALARM 2 WARN
H
2
S/TRS concentration alarm limit
#2 exceeded
WTRSALARM1
7+5
TRS ALARM 1 WARN
TRS concentration alarm limit #1
exceeded
WTRSALARM2
7+5
TRS ALARM 2 WARN
TRS concentration alarm limit #2
exceeded
WPMT
PMT DET WARNING
PMT detector outside of warning
limits specified by DETECTOR_LIMIT
variable.
WUVLAMP
UV LAMP WARNING
UV lamp reading outside of warning
limits specified by DETECTOR_LIMIT
variable.
WOXYFLOW
6
OXY FLOW WARNING
Oxygenator flow outside of warning
limits specified by OXY_FLOW_SET
variable.
WSAMPFLOW
SAMPLE FLOW WARN
Sample flow outside of warning
limits specified by SAMP_FLOW_SET
variable.
WSAMPPRESS
SAMPLE PRESS WARN
Sample pressure outside of warning
limits specified by SAMP_PRESS_SET
variable.
WCONVTEMP
4
CONV TEMP WARNING
Converter temperature outside of
warning limits specified by
CONV_SET
variable.
WBOXTEMP
BOX TEMP WARNING
Chassis temperature outside of
warning limits specified by BOX_SET
variable.
WRCELLTEMP
RCELL TEMP WARNING
Reaction cell temperature outside of
warning limits specified by
RCELL_SET
variable.
WIZSTEMP
IZS TEMP WARNING
IZS temperature outside of warning
limits specified by IZS_SET variable.
07266B DCN6485
Содержание T101
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank ...
Страница 10: ...vi This page intentionally left blank ...
Страница 12: ...viii This page intentionally left blank ...
Страница 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Страница 34: ...Getting Started Teledyne ML T101 Operation Manual 30 3 2 2 REAR PANEL Figure 3 4 Rear Panel Layout ...
Страница 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Страница 190: ...Calibration Procedures Teledyne ML T101 Operation Manual 186 This page intentionally left blank ...
Страница 202: ...Instrument Maintenance Teledyne ML T101 Operation Manual 198 047400102 Rev XC This page intentionally left blank ...
Страница 204: ...Instrument Maintenance Teledyne ML T101 Operation Manual 200 This page intentionally left blank ...
Страница 247: ...Teledyne ML T101 Operation Manual Troubleshooting Service 243 This page intentionally left blank ...
Страница 248: ......
Страница 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Страница 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Страница 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Страница 345: ...D 7 07266B DCN6485 ...