Teledyne ML – T101 Operation Manual
Remote Operation
145
Table 5-1. Status Output Pin Assignments
CONNECTOR PIN
STATUS
CONDITION (ON=CONDUCTING)
1
System Ok
ON if no faults are present.
2
Conc Valid
ON if concentration measurement is valid, OFF when invalid.
3
High Range
ON if unit is in high range of any AUTO range mode.
4
Zero Cal
ON whenever the instrument is in ZERO calibration mode.
5
Span Cal
ON whenever the instrument is in SPAN calibration mode.
6
Diag Mode
ON whenever the instrument is in DIAGNOSTIC mode.
7
Low Range
ON if unit is in low range of any AUTO range mode.
8
Unused
D
Emitter Bus
The emitters of the transistors on pins 1-8 are bussed
together. For most applications, this pin should be connected
to the circuit ground of the receiving device.
+
Dc Power
+ 5 VDC source, 30 mA maximum (combined rating with
Control Inputs)
Digital Ground The ground from the analyzer’s internal, 5/±15 VDC power
supply.
5.1.1.2.
Control Inputs
Control inputs allow the user to remotely initiate ZERO and SPAN calibration modes are
provided through a 10-pin connector labeled CONTROL IN on the analyzer’s rear panel.
These are opto-isolated, digital inputs that are activated when a 5 VDC signal from the
“U” pin is connected to the respective input pin.
Table 5-2. Control Input Pin Assignments
INPUT
STATUS
CONDITION WHEN ENABLED
A
External Zero Cal Zero calibration mode is activated. The mode field of the
display will read ZERO CAL R.
B
External Span Cal Span calibration mode is activated. The mode field of the
display will read SPAN CAL R.
C
External Low
Span Cal
Low span (mid-point) calibration mode is activated. The mode
field of the display will read LO CAL R.
D
Unused
E
Unused
F
Unused
Digital Ground
Provided to ground an external device (e.g., recorder).
U
DC Power For
Input Pull Ups
Input for +5 VDC required to activate inputs A - F. This voltage
can be taken from an external source or from the “+” pin.
+
In5v
Supply
Internal source of +5V which can be used to activate inputs
when connected to pin U.
Содержание T101
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank ...
Страница 10: ...vi This page intentionally left blank ...
Страница 12: ...viii This page intentionally left blank ...
Страница 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Страница 34: ...Getting Started Teledyne ML T101 Operation Manual 30 3 2 2 REAR PANEL Figure 3 4 Rear Panel Layout ...
Страница 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Страница 190: ...Calibration Procedures Teledyne ML T101 Operation Manual 186 This page intentionally left blank ...
Страница 202: ...Instrument Maintenance Teledyne ML T101 Operation Manual 198 047400102 Rev XC This page intentionally left blank ...
Страница 204: ...Instrument Maintenance Teledyne ML T101 Operation Manual 200 This page intentionally left blank ...
Страница 247: ...Teledyne ML T101 Operation Manual Troubleshooting Service 243 This page intentionally left blank ...
Страница 248: ......
Страница 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Страница 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Страница 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Страница 345: ...D 7 07266B DCN6485 ...