Operating Instructions
Teledyne ML – T101 Operation Manual
134
4.8.2.4.
Editing DAS Parameters
Data channels can be edited individually from the front panel without affecting other data
channels. However, when editing a data channel, such as during adding, deleting or
editing parameters, all data for that particular channel will be lost, because the DAS can
store only data of one format (number of parameter columns etc.) for any given channel.
In addition, an DAS configuration can only be uploaded remotely as an entire set of
channels. Hence, remote update of the DAS will always delete all current channels and
stored data.
To modify, add or delete a parameter, follow the instruction shown in Section 4.11.2.2
then press:
Edit Data Parameter Menu
Edit Data Channel Menu
SETUP X.X
0)
CONC: ATIMER, 2, 4032, R
PREV NEXT INS DEL
EDIT
PRNT
EXIT
Exits to the main
Data Acquisition
menu
SETUP X.X
NAME:CONC
<SET
SET>
EDIT
EXIT
SETUP X.X
PARAMETERS : 2
<SET SET>
EDIT
EXIT
SETUP X.X
0) PARAM=S2SCN1, MODE=AVG
PREV NEXT
INS DEL EDIT EXIT
Press
SET>
until…
SETUP X.X
EDIT PARAMS (DELETE DATA)
YES NO
NO returns to
the previous
menu and
retains all data.
Moves the
display between
available
Parameters
Inserts a new Parameter
before the currently
displayed Parameter
Deletes the Parameter
currently displayed.
Use to configure
the functions for
this Parameter.
Exits to the main
Data Acquisition
menu
YES will delete
all data in that
entire channel.
From the
DATA ACQUISITION
menu
Содержание T101
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank ...
Страница 10: ...vi This page intentionally left blank ...
Страница 12: ...viii This page intentionally left blank ...
Страница 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Страница 34: ...Getting Started Teledyne ML T101 Operation Manual 30 3 2 2 REAR PANEL Figure 3 4 Rear Panel Layout ...
Страница 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Страница 190: ...Calibration Procedures Teledyne ML T101 Operation Manual 186 This page intentionally left blank ...
Страница 202: ...Instrument Maintenance Teledyne ML T101 Operation Manual 198 047400102 Rev XC This page intentionally left blank ...
Страница 204: ...Instrument Maintenance Teledyne ML T101 Operation Manual 200 This page intentionally left blank ...
Страница 247: ...Teledyne ML T101 Operation Manual Troubleshooting Service 243 This page intentionally left blank ...
Страница 248: ......
Страница 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Страница 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Страница 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Страница 345: ...D 7 07266B DCN6485 ...