Teledyne ML – T101 Operation Manual
Operating Instructions
103
Once a TEST function is selected, the instrument begins to report a signal on the
A4
output and adds
TEST=
to the list of test functions viewable on the display (just before
the TIME display).
4.7.
SETUP – COMM: SETTING UP THE ANALYSER’S
COMMUNICATION PORTS
For remote operation the T101 is equipped with an Ethernet port, a USB port and two
serial communication (com) ports located on the rear panel. Both com ports (labeled
RS232, which is the COM1 port and COM2) operate similarly and give the user the
ability to communicate with, issue commands to, and receive data from the analyzer
through an external computer system or terminal. By default, both ports operate on the
RS-232 protocol.
The RS232 port can also be configured to operate in single or RS-232 Multidrop mode
(Option 62). The COM2 port can be left in its default configuration for standard RS-232
operation, or reconfigured for half-duplex RS-485 operation; (contact the factory for
configuration information). When COM2 is configured for RS-485 communication, the
rear panel USB port is disabled.
A code-activated switch (CAS), can also be used on either port to connect typically
between 2 and 16 send/receive instruments (host computer(s) printers, data loggers,
analyzers, monitors, calibrators, etc.) into one communications hub. Contact Teledyne
ML’s Sales for more information on CAS systems.
4.7.1.
INSTRUMENT ID
Each type of Teledyne ML analyzer is configured with a default ID code. The default ID
code for all T101 analyzers is either “
0
” or
101
. The ID number is only important if more
than one analyzer is connected to the same communications channel such as when several
analyzers are on the same Ethernet LAN, in an RS-232 Multidrop chain, or operating over
a RS-485 network. If two analyzers of the same model type are used on one channel, the
ID codes of one or both of the instruments needs to be changed so that they are unique to
the instruments.
Содержание T101
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank ...
Страница 10: ...vi This page intentionally left blank ...
Страница 12: ...viii This page intentionally left blank ...
Страница 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Страница 34: ...Getting Started Teledyne ML T101 Operation Manual 30 3 2 2 REAR PANEL Figure 3 4 Rear Panel Layout ...
Страница 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Страница 190: ...Calibration Procedures Teledyne ML T101 Operation Manual 186 This page intentionally left blank ...
Страница 202: ...Instrument Maintenance Teledyne ML T101 Operation Manual 198 047400102 Rev XC This page intentionally left blank ...
Страница 204: ...Instrument Maintenance Teledyne ML T101 Operation Manual 200 This page intentionally left blank ...
Страница 247: ...Teledyne ML T101 Operation Manual Troubleshooting Service 243 This page intentionally left blank ...
Страница 248: ......
Страница 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Страница 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Страница 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Страница 345: ...D 7 07266B DCN6485 ...