Teledyne ML - T100 UV Fluorescence SO2 Analyzer
Operating Instructions
97
4.6.7. RANGE UNITS
The T100 can display concentrations in parts per billion (10
9
mols per mol, PPB), parts
per million (10
6
mols per mol, PPM), micrograms per cubic meter (µg/m
3
, UGM) or
milligrams per cubic meter (mg/m
3
, MGM). Changing units affects all of the display,
analog outputs, COM port and DAS values for all reporting ranges regardless of the
analyzer’s range mode.
To change the concentration units:
SETUP X.X
CONC UNITS: PPM
PPM PPB
UGM
MGM
ENTER EXIT
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET
UNIT
EXIT
SETUP X.X
CONC UNITS: PPB
PPM
PPB
UGM
MGM
ENTER
EXIT
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
Select the preferred
concentration unit.
EXIT
returns
to the main menu.
ENTR
accepts
the new unit,
EXIT
returns
to the
SETUP
menu.
Figure 4-14:
SETUP RNGE
– Concentration Units Selection
IMPORTANT
Concentrations displayed in mg/m
3
and µg/m
3
use 0°C and 760 Torr as standard
temperature and pressure (STP). Consult your local regulations for the STP used by your
agency. Here are the conversion factors from volumetric to mass units used in the T100:
SO
2
: ppb x 1.34 = µg/m
3
; ppm x 1.34 = mg/m
3
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...