
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
Troubleshooting & Repair
06807A(DCNxxxx)
291
Figure 10-12: Disassembling the Shutter Assembly
6. Carefully remove the UV filter.
7. Install the UV filter.
8. Handle carefully and never touch
the filter’s surface.
9. UV
filter’s wider ring side should be facing out.
10. Install UV filter retainer and tighten screws.
11. Install the shutter cover and minifit connector. Tighten 4 shutter cover screws.
12. Reinstall the sensor module and Plug J4 connector into the motherboard.
10.7.2.5.
ADJUSTING THE UV LAMP (
PEAKING THE LAMP
)
There are three ways in which ambient conditions can affect the UV Lamp output and
therefore the accuracy of the SO
2
concentration measurement. These are:
Line Voltage Change:
UV lamp energy is directly proportional to the line voltage. This
can be avoided by installing adequate AC Line conditioning equipment such as a
UPS/surge suppressor.
Lamp Aging
- Over a period of months, the UV energy will show a downward trend and
can be up to 50% in the first 90 days, and then a slower rate, until the end of useful life of
the lamp. Periodically running the UV lamp calibration routine (refer to Section 4.8.7)
will compensate for this until the lamp output becomes too low to function at all.
Note
As the lamp degrades over time, the software for the CPU compensates for the loss of UV
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...