Troubleshooting & Repair
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
288
06807A(DCNxxxx)
10.7.2.3.
CLEANING THE PMT LENS AND PMT FILTER
IMPORTANT
The sample chamber should only be opened or cleaned on instructions from
the Teledyne ML Customer Service Department.
Be careful not to leave thumbprints on the interior of the sample chamber. The
various oils that make up fingerprints fluoresce brightly under UV light and will
significantly affect the accuracy of the analyzer’s SO
2
measurement)
To clean the PMT Lens and filter:
1. Remove the sensor module as described in Section 10.7.2.1.
Figure 10-9:
Hex Screw Between Lens Housing and Sample chamber
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...