Troubleshooting & Repair
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
268
06807A(DCNxxxx)
10.2.3. RELAY BOARD STATUS LEDS
The most important status LED on the relay board is the red I
2
C Bus watch-dog LED,
labeled
D1
(or W/D), which indicates the health of the I
2
C communications bus. This
LED is located in the upper left-hand corner of the relay board when looking at the
electronic components. If D1 is blinking, then the other LEDs can be used in conjunction
with the
DIAG
menu I/O functions to test hardware functionality by switching devices on
and off and watching the corresponding LED turn on or off. The LED only indicates that
the logic signal for an output has been activated. If the output driver (i.e. the relay or
valve driver IC) is defective, then the LED will light up, but the attached peripheral
device will not turn on.
Table 10-3: Relay Board Status LEDs
LED
COLOR
FUNCTION
FAULT
STATUS
INDICATED FAILURE(S)
D1
red
Watchdog Circuit;
I
2
C bus operation.
Continuously
ON or OFF
Failed/Halted CPU
Faulty Mother Board, Valve Driver board or
Relay PCA
Faulty Connectors/Wiring between Motherboard,
Valve Driver board or Relay PCA
Failed/5 VDC Power Supply (PS1)
10.3. GAS FLOW PROBLEMS
The standard analyzer has one main flow path. With the IZS option installed, there is a
second flow path through the IZS oven that runs whenever the IZS is on standby to purge
SO
2
from the oven chamber. The IZS flow is not measured so there is no reading for it on
the front panel display. The full flow diagrams of the standard configuration (refer to
Figure 3-18) and with options installed (refer to Figure 3-19 and Figure 3-20) help in
troubleshooting flow problems. In general, flow problems can be divided into three
categories:
Flow is too high
Flow is greater than zero, but is too low, and/or unstable
Flow is zero (no flow)
When troubleshooting flow problems, it is essential to confirm the actual flow rate
without relying on the analyzer’s flow display. The use of an independent, external flow
meter to perform a flow check as described in Section 10.5.2 is essential.
10.3.1. ZERO OR LOW SAMPLE FLOW
If the pump is operating but the unit reports a
XXXX
gas flow, do the following three
steps:
Check for actual sample flow
Check pressures
Carry out a leak check
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...