Theory Of Operation
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
252
06807A(DCNxxxx)
This measurement is stored in the analyzer. Memory as the test function
PMT TEMP
and
is viewable
as a test function (refer to Section 4.2.1) through the analyzer’s front
panel.
SAMPLE GAS PRESSURE SENSOR:
This sensor measures the gas pressure at the
exit of the sample chamber.
SAMPLE FLOW SENSOR:
This sensor measure the flow rate of the sample gas as it
exits the sample chamber.
9.5.9.3.
THERMISTOR INTERFACE
This circuit provides excitation, termination and signal selection for several negative-
coefficient, thermistor temperature sensors located inside the analyzer. They are as
follows:
SAMPLE CHAMBER TEMPERATURE SENSOR:
The source of this signal is a
thermistor imbedded in the of the sample chamber block. It measures the temperature of
the sample gas in the chamber. The data are used by the CPU to control sample chamber
the heating circuit and as part of the SO
2
, calculations when the instrument’s
Temperature/Pressure Compensation feature is enabled.
This measurement is stored in the analyzer. Memory as the Test Function
RCEL TEMP
and
is viewable
as a test function (refer to Section 4.2.1) in the analyzer’s front panel
display.
IZS OPTION PERMEATION TUBE TEMPERATURE SENSOR:
This thermistor,
attached to the permeation tube in the IZS option, reports the current temperature of that
tube to the CPU as part of control loop that keeps the tube at a constant temperature.
BOX TEMPERATURE SENSOR:
A thermistor is attached to the motherboard. It
measures the analyzer’s internal temperature. This information is stored by the CPU and
can be viewed by the user for troubleshooting purposes through the front panel display.
This measurement is stored in the analyzer‘s memory as the test function
BOX TEMP
and
is viewable
as a test function (refer to Section 4.2.1) in the analyzer’s front panel
display.
9.5.10. ANALOG OUTPUTS
The analyzer comes equipped with four Analog Outputs:
A1
,
A2
,
A3
and a fourth that is
a spare.
A1 and A2 Outputs:
The first two,
A1
and
A2
are normally set up to operate in parallel
so that the same data can be sent to two different recording devices. While the names
imply that one should be used for sending data to a chart recorder and the other for
interfacing with a data logger, either can be used for both applications.
Both of these channels output a signal that is proportional to the SO
2
concentration of
the sample gas. The
A1
and
A2
outputs can be slaved together or set up to operated
independently. A variety of scaling factors are available, refer to Section 4.6 for
information on setting the range type and Section 4.8.4 for adjusting the electronic
scaling factors of these output channels
Test Output:
The third analog output, labeled
A3
is special. It can be set by the user
(refer to Section 4.8.10) to carry the signal level of any one of the parameters accessible
through the
TEST
menu of the unit’s software.
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...