Calibration Procedures
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
196
6.10.2.4.
CO
2
ZERO/SPAN CALIBRATION
To perform the zero/span calibration procedure:
Press ENTR to changes
the OFFSET & SLOPE
values for the O
2
measurement.
Press EXIT to leave the
calibration unchanged and
return to the previous
menu.
SAMPLE
GAS TO CAL:CO2
SO2
CO2
O2
ENTR
EXIT
SAMPLE
CO2 STB=X.XX %
CO2=XX.XX
< TST TST > CAL SETUP
M-P CAL
CO2 STB=X.XX %
CO2=XX.XX
<TST TST>
ZERO
CONC
EXIT
EXIT at this point
returns to the
SAMPLE menu.
Press ENTR to changes
the OFFSET & SLOPE
values for the CO
2
measurement.
Press EXIT to leave the
calibration unchanged and
return to the previous
menu.
Allow zero gas to enter the sample port
at the rear of the analyzer.
Wait until CO2 STB
falls below 0.01%.
This may take several
minutes.
SAMPLE
CO2 STB=X.XX %
CO2=XX.XX
< TST TST >
CAL
SETUP
M-P CAL
CO2 STB=X.XX %
CO2=XX.XX
<TST TST>
ENTR
CONC
EXIT
SAMPLE
GAS TO CAL:CO2
SO2
CO2
O2
ENTR
EXIT
M-P CAL
CO2 STB=X.XX %
CO2=XX.XX
<TST TST> ZERO
SPAN
CONC
EXIT
Allow span gas to enter the sample port
at the rear of the analyzer.
Wait until CO2 STB
falls below 0.01%
This may take several
minutes.
SAMPLE
CO2 STB=X.XX
CO2=XX.XX
< TST TST >
CAL
SETUP
M-P CAL
CO2 STB=X.XX %
CO2=XX.XX
<TST TST>
ENTR
CONC
EXIT
M-P CAL
CO2 STB=X.XX %
CO2=XX.XX
<TST TST> ENTR
CONC
EXIT
The SPAN key now appears
during the transition from
zero to span.
You may see both keys.
If either the ZERO or SPAN
buttons fail to appear see the
chapter on Troubleshooting
for tips.
SAMPLE RANGE=50.0 PPM CO2= XX.XX
< TST
TST >
CAL SETUP
Set the Display to show
the CO2 STB test
function.
This function calculates
the stability of the CO
measurement.
Toggle TST> button until ...
Press CO2: ENTR to
initiate zero point
calibration of the
O2 sensor.
Press CO2: ENTR to
initiate span point
calibration of the
O2 sensor.
Figure 6-20:
CO
2
Zero/Span Calibration
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...