Calibration Procedures
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
186
IMPORTANT
The CALIBRATE attribute must always be set to OFF for analyzers used in US
EPA controlled applications that have IZS option installed.
Calibration of instruments used in US EPA related applications should only be
performed using external sources of zero air and span gas with an accuracy
traceable to EPA or NIST standards and supplied through
the analyzer’s sample
port (refer to Section 6.2).
The following example sets Sequence 2 to carry out a zero-span calibration every other
day starting at 01:00 on September 4, 2002, lasting 15 minutes. This sequence will start
0.5 hours later each day.
Table 6-4: Example Auto-Cal Sequence
MODE / ATTRIBUTE
VALUE
COMMENT
SEQUENCE
2
Define Sequence #2
MODE
ZERO-SPAN
Select Zero and Span Mode
TIMER ENABLE
ON
Enable the timer
STARTING DATE
Sept. 4, 2002
Start after Sept 4, 2002
STARTING TIME
01:00
First Span starts at 01:00
DELTA DAYS
2
Do Sequence #2 every other day
DELTA TIME
00:30
Do Sequence #2 0.5 h later each day
DURATION
15.0
Operate Span valve for 15 min
CALIBRATE
ON
The instrument will re-set the slope and offset
values for the SO
2
channel at the end of the AutoCal
sequence
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...