October 2017
RiverPro and RioPro ADCP Guide
Page 86
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
OL – Display Feature List
Purpose
Lists the special firmware upgrades that are installed.
Format
OL
Recommended Setting. Use as needed.
Description
Lists special features that are installed. See
Feature Upgrades
for information on how to
install additional capability in the RiverPro/RioPro.
Examples
See below.
>ol FEATURES
---------------------------------------------------------------
Feature Installed
---------------------------------------------------------------
Manual Profile Modes Yes
Suspended Sediment No
SxS No
See your technical manual or contact TRDI for information on
how to install additional capability in your RiverPro/RioPro.
>
The
Section-By-Section (SxS)
feature allows transects to be completed by measuring at
different user selected points on the river. This makes it possible to measure streams that are
frozen over.
Manual Profile Modes allow the RiverPro/RioPro to override the automatic profiling mode
and adds additional water profiling commands (see
Contact your local sales representative if you are interested in upgrading your system.
Y – Display Banner
Purpose
Displays the RiverPro/RioPro banner.
Format
Y
Recommended Setting.
Use as needed
.
Description
Example
Y
>y
RiverPro
Teledyne RD Instruments (c) 2015
All rights reserved.
Firmware Version: 56.xx