3.5.3. Calibration Logs
For each calibration, the monitor logs the following:
• Calibration TIME and DATE
• Calibration method (ABC)
• Before Zero gas reading for % Volume Gas and ppm CO
• After Zero gas reading for % Volume Gas and ppm CO
• Before Span gas reading for % Volume Gas and ppm CO
• After Span gas reading for % Volume Gas and ppm CO
3.5.4. Datalogging Capacity/date Format
• Timed logs:24 hours at 1 log per minute
• Session logs:Minimum of 180
• Calibration logs:Minimum of 8
If the datalogging memory is full, the monitor prompts the user. Also:
• User
CANNOT
enter INSPT mode
• User can still enter PS and PURGE modes
3.5.5. Datalogging Extraction
Results from leak surveys (INSPT Mode) can be transferred via Bluetooth to applications running
on tablets, allowing the data to be visualized and then stored to the cloud for regulatory reporting
purposes by a Utility.
3.6. View Maximum Recorded Values Since Switch
ON
The monitor records the maximum values for each sensor since switch on. To view the maximum
values in any mode:
1. From the normal operating display, press the R button once to switch the monitor backlight ON
2. Press the R button again, while the screen light is ON, to view the stored maximum % VOL
reading (see
). Note that MAX is displayed at the
bottom center of the display.