Appendix G: Spectrograph Adapters
A PI-MAX3 with the spectroscopy-mount adapter can be readily mounted to a Teledyne
Acton Research Series spectrograph or to an IsoPlane SCT-320 spectrograph. Mounting
the camera to a Teledyne Acton Research Series spectrograph requires a male sliding
tube that slides into the spectrograph for setting the focus. Mounting a PI-MAX3 to an
IsoPlane is a matter of bolting it directly to the mounting plant on the IsoPlane.
NOTES:
1. When mounted to a spectrograph, the text on the back
of the PI-MAX3 should be right side up.
2. A C-mount adapter plate and nut are available for
mounting a camera with a C-mount adapter to a
Teledyne Acton Research Series spectrograph. Refer to
Section G.3, PI-MAX3 to Teledyne Acton Research
(C-Mount Adapter)
, on page 233 for mounting
instructions.
G.1
Spectrograph-Detector Focus
The detector mounting hardware provides two degrees of freedom: focus and rotation.
In this context, focus means to physically move the detector back and forth through the
focal plane of the spectrograph. The approach taken is to slowly move the detector in
and out of focus and adjusting for optimum while watching a live display on the
monitor, followed by rotating the detector and again adjusting for optimum. The
following procedure, which describes the focusing operation with a Teledyne Acton
Research SP-2300i spectrograph, can be easily adapted to other spectrographs. For
IsoPlane SCT-320 related focusing information, refer to
NOTES:
1. Use a USB-powered or a DC-powered light source. The
60 Hz of an AC-powered light source may make it more
difficult to achieve focus.
2. If you must use an AC light source, set the
SuperSYNCHRO Internal frequency to around 10 Hz.
G.1.1
Teledyne Acton Research Series Spectrograph
Perform the following procedure to focus a PI-MAX3 when mounted on a Teledyne
Acton Research series spectrograph:
1. Mount a light source such as a mercury pen-ray type in front of the entrance slit of
the spectrograph. Any light source with line output can be used. Standard
fluorescent overhead lamps have good calibration lines as well. If there are no line
sources available, it is possible to use a broad band source such as tungsten for the
alignment. If this case, use a wavelength setting of 0.0 nm for alignment purposes.
Содержание Princeton Instruments PI-MAX 3
Страница 1: ...www princetoninstruments com PI MAX 3 System Manual 4411 0129 Issue 2 August 20 2019 ...
Страница 32: ...32 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 38: ...38 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 124: ...124 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 150: ...150 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 174: ...174 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 178: ...178 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 190: ...190 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 214: ...214 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 236: ...236 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 244: ...244 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 249: ...This page is intentionally blank ...