16.1 Alarm Sounds Repetitively
Immediately reduce the light entering the camera. This can be done by switching the
I.I.T. switch on the back of the PI-MAX3 to OFF until you lower the source illumination
by either decreasing the lens aperture or by completely blocking the light into the
camera window with a lens cap (or equivalent) until the light level has been lowered.
If the alarm sounds repetitively even when the illumination level is adequately low,
switch the I.I.T. to OFF and turn off the PI-MAX3 power supply. Then contact the
factory. This may indicate intensifier damage or another situation that requires
immediate attention.
16.2 Alarm Sounds Sporadically
It is normal for the alarm to sound briefly when the high-voltage supply is first turned
on. However, if the alarm sounds sporadically, contact the factory at once. This may
indicate intensifier damage or another situation that requires immediate attention.
Refer to
on page 248 for complete information.
16.3 Baseline Signal Suddenly Changes by > 1000 ADU
There are two possible reasons for this change:
•
The temperature setting has been changed.
In this case, a change in baseline signal is normal.
•
There may be excessive humidity in the CCD enclosure of the camera.
If the temperature setting has not been changed and you observe a baseline
signal change, turn off the system immediately. An excess humidity condition
should be corrected promptly or permanent damage not covered by the
Warranty could occur. Have the unit serviced by Teledyne Princeton
Instruments or an authorized Teledyne Princeton Instruments service facility.
16.4 Camera Is Not Responding
If this message pops up when you click OK after selecting the Interface Type during
Hardware Setup (under the WinX/32 Setup menu,) the system has not been able to
communicate with the camera. Verify the Controller has been turned ON and that the
interface card, its driver, and the interface cable have been installed.
•
If the Camera is ON, the problem may be with the interface card, its driver,
interrupt or address conflicts, or the cable connections.
•
If the interface card is not installed, close the application software and turn the
PI-MAX3 power supply OFF.
Follow the interface card manufacturer's installation instructions and cable the
interface card to the GIG-E port on the rear of the camera. Then perform a
Custom installation of WinSpec/32 or WinView/32 with the appropriate
interface component selected. Be sure to deselect the interface component
that does not apply to your system.
•
If the interface card is installed in the computer and is cabled to the GIG-E port
on the rear of the camera, close the application program and turn the camera
OFF. Check the cable connections and secure the cables if connections are
loose.
•
If the interface card was installed after the WinX/32 application software has
been installed, close the application program and do a Custom installation of
WinSpec/32 or WinView/32 with the GigE interface component selected. Be
sure to deselect the interface component that does not apply to your system.
Содержание Princeton Instruments PI-MAX 3
Страница 1: ...www princetoninstruments com PI MAX 3 System Manual 4411 0129 Issue 2 August 20 2019 ...
Страница 32: ...32 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 38: ...38 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 124: ...124 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 150: ...150 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 174: ...174 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 178: ...178 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 190: ...190 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 214: ...214 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 236: ...236 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 244: ...244 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 249: ...This page is intentionally blank ...