183
Chapter 15: Tips and Tricks
,
Chapter 8, WinX/32 and Gated Operation
LightField and Gated Operation
, the objective was to show how to take an image or
spectrum with a minimal operating configuration and following a simple procedure. In
this chapter, we consider factors that affect more complex measurements.
15.1 Overexposure Protection
WARNING!
Image intensified detectors such as the PI-MAX3 can be
destroyed if exposed to excessive light levels. Teledyne
Princeton Instruments cannot take responsibility for
damage due to misuse.
Intensified detectors must not be continuously exposed to high-level radiation (
≥
10
-4
foot candles). When the illumination level is not quantitatively known, toggle the I.I.T.
switch (on the back of the PI-MAX3) to the OFF position while you are adjusting the
incoming light level. After making adjustments, toggle the switch to the ON position. If
the alarm sounds repetitively, toggle the switch back to OFF and readjust the lighting
conditions.
If the experimental conditions dictate that only a small portion of the photocathode is
illuminated over relatively long periods of time, change the illuminated region of the
photocathode periodically to avoid long term localized photocathode or MCP damage.
If you can do so, it is a good idea to avoid intense light on the intensifier even when it is
off. High light levels increase EBI, often for hours, even when the intensifier is off. Image
burn-in can occur with some photocathode types.
NOTE:
The audible alarm and protection circuits are not fail-safe
protection, particularly when working with high intensity
sources such as lasers. For additional information, refer to
15.2 Signal Delay
For the detector to see a transient signal, it is essential that it be gated on when the
signal arrives at the detector. If this requirement isn't given careful consideration, it is
possible to set up an experiment in which the signal will come and go before the
detector is gated on. If this happens, no proper data can be taken. Depending on the
nature of the experiment and the specific equipment involved, a number of different
factors may need to be considered to be assured that the detector gates at the right
time.
Содержание Princeton Instruments PI-MAX 3
Страница 1: ...www princetoninstruments com PI MAX 3 System Manual 4411 0129 Issue 2 August 20 2019 ...
Страница 32: ...32 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 38: ...38 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 124: ...124 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 150: ...150 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 174: ...174 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 178: ...178 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 190: ...190 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 214: ...214 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 236: ...236 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 244: ...244 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 249: ...This page is intentionally blank ...