Periodic Maintenance Items
These maintenance items should be done prior to using the DVL.
Cleaning the Cable Connectors
After a deployment, clean and remove any accumulated sand or mud from the both the connector on the
Pathfinder DVL and the socket on the cable.
To clean the connectors:
1. Flush the connector pins and sockets with soap and fresh water (deionized water if available) to
remove all dirt, grit, and lubricant.
2. Rinse out with alcohol, allowing the connector to air dry.
Do
NOT
use spray-based contact cleaner. The use of some oil-based propellants in spray cans
can cause conductivity problems in neoprene.