Pathfinder DVL Guide
April
2018
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
Page 23
Testing the Pathfinder System
To test the Pathfinder:
1. Interconnect the system as described in Setting Up the Pathfinder System.
2. Place the Pathfinder transducer in water (at least a few inches to cover the transducer face).
3. Apply power to the system.
4. Start the TRDI Toolz program (for help on using TRDI Toolz, see the TRDI Toolz Help file).
5. Click the Break button. The wakeup banner appears on the log file window.
Pathfinder
Teledyne RD Instruments (c) 2015
All rights reserved.
Firmware Version: 67.xx
>
6. At the “>” prompt in the communication window, enter the command
CR1
then press the Enter
key. This will set the Pathfinder to the factory default settings.
7. At the “>” prompt in the communication window, enter the command
CK
then press the Enter
key. This will save the factory default setting.
8. At the “>” prompt in the communication window, enter the command
PS0
then press the Enter
key. This will display the Pathfinder system configuration data.
>ps0
Serial Number: 0
Frequency: 614400 Hz
Configuration: DVL : 4-beam velocity.
Transducer Type: PHASEDARRAY
Beam Angle: 30 Degrees
Beam Pattern: CONVEX
Sensors: TEMP
CPU Firmware: 67.00c MH0
FPGA Version: 1.00.003 [0x1003]
Board Serial Number Data:
FF 00 00 08 0A 1A E1 28 DS18B20 TMP SNS
52 00 00 02 7B 85 3C 23 ETH72B-2305-00A
5F 00 00 02 61 E5 71 23 DSP72B-2303-00A
B5 00 00 02 1A 90 C3 23 PTX72B-2304-00A
C3 00 00 02 63 57 0B 23 RCV72B-2302-03A
C6 00 00 02 7B 8C F7 23 XDC72B-1110-00B
9. At the “>” prompt in the communication window, enter the command
PA
then press the Enter
key. This will run the Pathfinder Pre-Deployment test.
Submerge the transducer face in a few inches of water.
The PA test may fail in air.
Only a
failure with the transducer in water is a relevant test.
This test may fail if the Pathfinder is exposed to electrical interference such as that from
other acoustic devices or SONARs or other electronic devices (i.e. cellphones, radios,
computers, TV’s, etc.). The test may also fail if the Pathfinder or its cable is exposed or
installed near potential sources of EMI’s such as high voltage lines like main engine cables,
galley equipment, winches, engine room equipment, cranes, high voltage lighting circuit, etc.
Interference from the mentioned sources may contaminate the Pathfinder’s data as well.
Oftentimes, a noisy (i.e. non-UPS) power supply may cause failures in this test as well.