Pathfinder DVL Guide
Page 134
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
PC5 or PC 50
The PC5 test updates the data inline using cursor commands, and the PC50 test performs the same test,
but output the data in a line-by-line fashion to accommodate terminals that do not properly support cur-
sor commands.
>pc50
Leak detector ADC data is sampled and displayed in a loop.
Press any key to exit the loop.
DetA DetB
0dc0 0008
0dc1 0007
0dc2 0004
>
The leak sensor status can be decoded as follow:
1-
Convert hex asci to decimal. For example:
a.
DetA(hex) 0dc0 → to dec = 3520
b.
DetB(hex) 0008 → to dec = 8
2-
Convert these values to Voltages by x2.5 and /4095. For example:
a.
DetA(hex) 3520*2.5/4095= 2.14
b.
DetB(hex) 0008*2.5/4095= 0.0048
c.
Compare the Values to the threshold shown below to determine leak
status:
Leak
No Leak
Sensor Not Connected
Less than 1.24 Volts
1.25 to 2.3 Volts
Greater than 2.3 Volts
In our example above, Leak Sensor A is operating properly and did not detect a leak while
Leak Sensor B
is detecting a leak
.
PS – Display System Parameters
Purpose
Displays Pathfinder system configuration data.
Format
PS
n
Range
n
= 0, 1, 3 to 5
PS0 – System Configuration Info
>ps0
Serial Number: 3002
Frequency: 307200 Hz
Configuration: DVL : 4-beam velocity.
Transducer Type: PHASEDARRAY
Beamformer Rev: XA0
Beam Angle: 30 Degrees
Beam Pattern: CONVEX
Sensors: AHRS TEMP PRESS 0BAR
CPU Firmware: 67.01
FPGA Version: 7.00.002 [0x7002]
Board Serial Number Data:
42 00 00 05 F5 CD 7A 28 DS18B20 TMP SNS
6C 00 00 01 E1 CA 84 23 RCV72B-2243-14E
FB 00 00 01 E8 4E EA 23 PIO72B-2301-03A
BA 00 00 01 D9 F6 1D 23 XDR72B-1001-00A
C0 00 00 01 DA 29 7D 23 DSP72B-2212-00C
34 00 00 01 DA 18 87 23 PER72B-2353-01A