Platinum/Optima CVE Refrigerated Sampler
3-12
sampling without interruption even if that
circuit reopens.
*Float Start/Stop- If the FLT S
TART
/S
TOP
setting is selected, the program will reset its
sampling interval counters and not restart
them until the float circuit closes again (i.e.,
those counters will restart from zero when
the water level rises high enough to close
the float switch).
Note
The full bottle switch uses the float start/stop feature. If
float start or start/stop is being used as a remote start
the full bottle switch cannot be used.
3.3.8 Refrigeration
The sampler will display the interior cabinet
temperature. Scroll thru the run displays to
locate the temperature.
Setting the Refrigerator Temperature:
The
thermostat is located on the lower right side of
the refrigeration module. Remove the air vent
to adjust the temperature by using the up and
down arrow on the display. To set the
temperature output 1 for refrigeration and
output 2 for the heater.