089770000 Rev 05
Teledyne API N500 CAPS NO
X
Analyzer with NumaView™ Software
81
The instrument’s electronic systems will operate with any of the specified power within
the 100 VAC to 240 VAC, at 47 Hz to 63 Hz. Using the properly rated power cord, it will
power on when the rear panel Hard Power switch is placed in the ON position. The status
LED located on the front panel Soft Power switch should begin blinking after the power
is supplied. If that status LED is not blinking:
•
Check the power cord for damage, such as whether it’s cut or burned.
•
Check that the power cord is adequately rated for the instrument’s specified power
rating.
•
Check that the power source is of the proper voltage for the instrument’s specified
power rating.
•
If there are no findings in the preceding steps, then note whether the instrument had
been opened for maintenance; if so, place the rear panel Hard Power switch in the
OFF position, and disconnect the power cord; then reopen the instrument and
check that no wiring had been dislodged, and no tools were left inside.
•
If no other reason can be found for the instrument not powering on, then check the
fuse with an ohmmeter to determine its viability: carefully follow the instructions in
Section 5.7.7.1 to remove the fuse for testing.
•
If the fuse is blown, replace it with a fuse of the correct specifications as
instructed in Section 5.7.7.1.
•
If the fuse is not blown, or if the replacement fuse blows, then call Technical
Support (Section 5.9).
LCD/D
ISPLAY
M
ODULE
Assuming that there are no wiring problems and that the DC power supplies are operating
properly, the display screen should light and show the splash screen and other indications
of its state as the CPU goes through its initialization process.
O
3
G
ENERATOR
M
ODULE
The ozone generator can fail in two ways, electronically (printed circuit board) and
functionally (internal generator components). Assuming that air is supplied properly to
the generator, the generator should automatically turn on 30 minutes after the instrument
is powered up. However, if the generator appears to be working properly but the
sensitivity or calibration of the instrument is reduced, suspect a leak in the ozone
generator supply air.
A leak in the dryer or between the dryer and the generator can reduce sensitivity and
cause performance drift. Carry out a leak check (Section 5.6.5).