4. Remove the AREF filter assembly from the mounting brackets.
5. Remove the fittings that connect the two filters to one another.
6. Install replacement filters and reconnect AREF assembly and fittings in reverse
order.
Figure 5-6. Pneumatic Layout of AREF Filter and Charcoal Filter
REPLACING THE INTERNAL PUMP
To replace the internal pump:
1. Turn OFF the analyzer (this is important to avoid potential for contamination
of the measurement cell).
2. Remove the top cover of the analyzer.
3. Locate the vacuum pump and using snips, remove the zip ties that secure
the four hoses that connect to the pump (two on each side)
(It may be a good idea to mark the tubes with color codes to be sure they are
reconnected properly).
4. Remove the four tubes from the hose barb connections on the pump.
5. Remove the four screws that hold the pump/bracket assembly to the bottom
of the chassis.
6. Disconnect the power connector labeled “pump” located about 4 inches
down the black/red cable coming from the top of the pump.
7. Pull pump assembly out and set aside.
8. Install replacement pump/bracket assembly
9. Be sure to reconnect tubes in the proper orientation and zip tie to secure
10. Replace screws for the pump mounting bracket into the bottom of the
chassis.
11. Connect the power connector.