58
Teledyne API N500 CAPS NO
X
Analyzer with NumaView™ Software
089770000 Rev 05
Figure 4-1. Multi-Point Calibration Page
Z
ERO
C
ALIBRATION
C
HECK AND
A
CTUAL
C
ALIBRATION
Go to the Calibration>M-P menu.
Input Zero air through the Sample port and press the Start button.
Either check or calibrate as follows:
CHECK ONLY:
ACTUAL CALIBRATION:
a. Wait for reading to stabilize.
b. Press Stop and check the
reading.
a. Press the Zero button.
b. Press Stop and check the reading.
S
PAN
C
ALIBRATION
C
HECK AND
A
CTUAL
C
ALIBRATION
1. While still in the Calibration>M-P menu, input NO
2
Span gas through the
Sample port and press the Start button.
2. Either check or calibrate as follows:
CHECK ONLY:
ACTUAL CALIBRATION:
a. Wait to reach stability,
then press Stop.
b. Record the reading(s).
a. Press the Set Span Target button and
enter the NO
2
and NOx concentrations.
b. Verify the concentration reading is the
same as the NO
2
concentration being
supplied.
c. If correct, wait to reach stability, then
press the Span button.
d. In the Cal Result window, press OK.
3. Press the Stop button and return to Home screen.
4. In the Dashboard, check and record the Slope(s) and the Offset(s).