6 – Maintenance
6-4
© Teledyne Limited
DPN 402197 Issue 7.1
6.3.1 Fault on a Single Channel
Figure 6-2: Single channel failure
Contact Teledyne
TSS
Power off the
system
Swap out the coil
cable
Is the channel
working?
Renew the faulty
coil cable
Swap out the coil on
the faulty channel
Is the channel
working?
Renew the faulty
coil
Swap out the SEP
Is the channel
working?
Renew the faulty
SEP
Single channel fault
Yes
No
Yes
No
Yes
No