HDO9000 High Definition Oscilloscopes Operator's Manual
maximum single-shot sampling rate. However, the real-time interval over which the instrument
collects the waveform data is much longer, and depends on the trigger rate and the amount of
interleaving required.
Interleaving of samples in RIS sampling mode.
Note:
RIS is not available when the oscilloscope is operating in fixed sample rate or HD Mode.
Sequence Sampling Mode
In Sequence Mode sampling, the completed waveform consists of a number of fixed-size segments.
The instrument uses the Timebase Sequence settings to determine the capture duration of each
segment. The desired number of segments, maximum segment length, and total available memory
are used to determine the actual number of samples or segments, and time or points.
Sequence Mode is ideal when capturing many fast pulses in quick succession or when capturing few
events separated by long time periods. The instrument can capture complicated sequences of events
over large time intervals in fine detail, while ignoring the uninteresting periods between the events.
Measurements can be made on selected segments using the full precision of the timebase.
Note:
Sequence Mode sampling is not available when operating in HD Mode.
Sequence Mode Set Up
The Sequence dialog appears only when Sequence Mode sampling is selected. Use it to define the
number of fixed-size segments to be acquired in single-shot mode.
1. From the menu bar, choose
Timebase > Horizontal Setup...
, then
Sequence Sampling Mode
.
2. On the
Sequence tab
under Acquisition Settings, touch
Number of Segments
and enter a value.
52