Param(eter) Compare Qualifiers
Pass/Fail qualifiers (Q
x
) can be configured to compare different parameter measurements (P
x
) to
each other or to a user-defined limit (or statistical range).
Comparing a Measurement to a Limit
This method takes the result of a selected measurement parameter and compares it to a user-
defined value or range of values.
1. Choose
Analysis > Pass/Fail
to display the
Pass/Fail
dialog.
2. Touch the
Q
x
button
where you want to set up a qualifier.
3. From the pop-up menu, choose
Param compare
.
4. On the Q
x
dialog, touch
Source1
and choose the parameter.
5. On the ParamCompare dialog, touch
Compare Values
and select
All
or
Any
from the pop-up.
By selecting
All
, the condition is true only if every waveform falls within the limit. Selecting
Any
makes the condition true if just one waveform falls within the limit.
6. Touch
Condition
and select a math operator, then touch
Limit
and enter the value that
completes the condition. The value entered acquires properties corresponding to the
parameters being tested. For example, if you are testing a time parameter, the unit is seconds.
7. If you chose to set a Delta limit, also enter the
Absolute or % Delta
value. You may choose
instead to set Limit and Delta using one of the buttons at the bottom of the dialog.
Comparing Parameter Measurements
When using Dual Param Compare, two measurements are compared, rather than two waveforms.
1. Choose
Analysis > Pass/Fail
to display the
Pass/Fail
dialog.
2. Touch the
Q
x
button
where you want to set up a qualifier.
3. From the Pass/Fail Condition menu, choose
Dual Param Compare
.
4. On the left-hand Qx dialog, select the parameters to compare in
Source1
and
Source2
.
5. On the ParamCompare dialog, touch
Compare values
and select
All
or
Any
from the pop-up.
By selecting
All
, the test are true only if every waveform falls within the set limit. Selecting
Any
makes the test true if just one waveform falls within the limit.
6. Touch
Condition
and select the math operator that completes the condition. The Summary
field shows the completed condition that is represented by this qualifier.