Figure 81: Bump Test - Purge
Note: The following gas concentrations are suggested if no other requirements are
specified by local regulations and/or internal company procedures:
•
Flammable ppm
•
Flammable %LEL
•
Flammable Volume
•
Oxygen
•
Carbon Monoxide (CO)
•
Hydrogen Sulphide (H
2
S)
4550ppm
50% LEL
100% Volume
18% Volume
500ppm
50ppm
7.3. Automatic Bump Test
For details on automatic bump testing using the GDUnet station, please refer to the following
documents:
•
GDUnet User Handbook
- part no. 14765
•
FlexiCal Plus User Handbook
- part no. 99551
Figure 82: GDUnet Station