Operator’s Manual
9
GAS LASER
Doc No. 16162 Rev A, GAS LASER Operator Manual
© 2021 Teledyne Gas Measurement Instruments Ltd. All rights reserved.
Specifications
General
Gas Laser Weight
3 lbs (approx.)
Carry Case Dimensions
21” x 17.5” x 9.5”
Display
3.5” color LCD
Storage
Internal SD card (not removable)
Power
Battery
Removable
Rechargeable
Lithium-ion pack, 10.8 VDC 3.2Ah
Battery Run Time
8 hours at 32° F (approx.)
Battery Charger
External
110-240 VAC, 50/60 Hz Universal
Charge Time
2-3 hours full charge (approx.)
Charging Indicator
Integrated into Dual Battery Charger
Detection/Measurement System
Detection Method
Tunable Diode Laser Absorption
Spectroscopy (TDLAS)
Detection Distance
100 ft (30m) nominal - may vary
due to background type and
conditions
Measurement Range
0 to 50K PPM-M
Sensitivity
5 PPM-M at distances from 0 to
100 ft (30m)
Beam Size
Conical in shape with a 22”
diameter at 100 ft (55cm at 30m)
Display
Resolution
320x240
Camera
Color
Aperture
f/2.6
FOV
94DEG (at 6.0mm image circle)
Lasers
IR Laser
Class I
Spotter Laser
On time duration is 2 minutes
Class 2 (II) <2mW @ 532nm
Spot size is 7mm at 15M
Complies with 21 CFR 1040.10
and 1040.11 except for deviations
pursuant to Laser Notice No. 50,
dated June 24, 2007
Eye Safety Warning
Do not stare into beam or
view directly with optical
instrument
Avertissement de sécurité des yeux
Ne jamais fixer le faisceau ni le
regarder directement avec des
instruments optiques.
GPS
Compatible With
GPS
GLONASS
Beidou
Galilieo
Communication
Bluetooth 4.2 BLE
(to support future features and
mobile applications)
WiFi
USB Dual Mode
USB Port Max Rating:
UM = 5V, IN = A
Alarms
Digital Methane Detection (DMD)
Audible tone and visible color
border when detection threshold
exceeded
Adjustable Detection Alarm Level
50’
1 to 200 PPM M
100’
1 to 400 PPM M
System Fault & Warnings
Audible alarm and visual
indication on the display
Testing
Built-In Self Test
Verifies operation and adjusts
laser wavelength for maximum
sensitivity
Test Gas Cell
Integrated within carrying case
Data Logging
Saves to Internal Memory
FAULT logs
Self Test logs
Captures
Data Collected
Includes, but not limited to:
CH4 PPM-M measurement
GPS location
Timestamp
Battery level
Battery voltage
Serial number of the instrument
Operating Conditions
Operating Temperature
0° to +122° F (-17° to 50° C)
Humidity
5 to 95% RH, non-condensing
Altitude
Up to 6560 ft (2000 m)
Environment of Use
Pollution degree 2 or better
Outdoor use
Regulatory
Instrument Protection
IP54 (water splash and dust
resistant)
Compliance
EMC (EN61000-6-2, EN6100-6-4)
Low Voltage Directive
(2014/35/EU)
Radio Equipment Directive
(2014/53/EU)
ETSI EN 301 489-1 v2.2.0
EN 61326-1:2013
47 CFR Part 15 & ICES-003
Ordinary Location Safety
UL 61010-1
CAN/CSA-C22.2 No 61010-1-12
Hazardous Location Safety
Class I, Zone 2, AEx ic op is IIA T4 Gc
Class I, Division 2, Group D
Intrinsically Safe