A-1
TYPICAL OPERATING PARAMETERS
Typical operating parameters are as follows:
Gas
Range
Resolution Zero
Accuracy
EXT
0 - 1000 ppm
1 ppm
5%
Typically 20%
1000 - 4500 ppm
10 ppm
10 - 20 %LEL
1%
PPM
0 - 1000 ppm
5 ppm
5%
Typically 20%
1000 - 4500 ppm
50 ppm
LEL
0 - 10%
0.1%
+/- 0.5%
2% +/- 1% LEL
10 - 100%
1%
Volume
0 - 5%
0.1%
+/- 2%
1% +/- 1% Gas
Gas
5% - 100%
1%
Carbon
0 - 1000 ppm
1 ppm +/- 15 ppm +/-15 ppm + 5% reading
Monoxide
Hydrogen 0 - 100 ppm
1 ppm
+/- 2 ppm +/-2 ppm + 10% reading
Sulphide
Oxygen
0 - 21%
0.1%
+/- 0.5%
+/- 0.5% + 3% reading
21 - 25%
1%
N/A
+/- 0.5% + 3% reading
Notes:
All the values above are at normal temperature and pressure.
Humidity is between 0% and 95% RH (non-condensing).
Pressure changes at the inlet and exhaust are minimised as they
may cause transient changes in reading.
A