Falcon GC
Maintenance & Troubleshooting
Teledyne Analytical Instruments
25
if the flame is lit by holding a reflective surface over
the FID exhaust and examining for condensation.
Check with a flow meter to ensure that the Air (ignite)
Pressure, the Air (Run) Pressure and the Hydrogen
Pressure values are consistent with those found in your
Data Package shipped with the instrument.
5.2 Adjusting FID Pressures
To adjust the pressure of in the FID please follow the steps described
below.
Note: For Ambient and N
2
Carrier Gas applications, the settings
below may be different. Refer to the Tech Note provided
with the specific instrument.
1.
Turn off the H
2
and air flow to the FID. If carrier gas is N
2
adjust
the flow to 5psi.
2.
Click “Remote” on the Chromperfect screen.
3.
Select the “Pressures” tab.
Содержание FALCON GC
Страница 10: ...Falcon GC x Teledyne Analytical Instruments Blank Page ...
Страница 15: ...Falcon GC Preparatio n Teledyne Analytical Instruments 5 3 Click on Report ...
Страница 16: ...Preparation Falcon GC 6 Teledyne Analytical Instruments Blank Page ...
Страница 42: ...Maintenance Troubleshooting Falcon GC 32 Teledyne Analytical Instruments ...
Страница 58: ...Maintenance Troubleshooting Falcon GC 48 Teledyne Analytical Instruments Figure 5 1 Glow Plug Wire ...
Страница 60: ...Maintenance Troubleshooting Falcon GC 50 Teledyne Analytical Instruments Figure 5 2 FID Diagram ...
Страница 62: ...Appendix Falcon GC 52 Teledyne Analytical Instruments A 3 Drawings ...
Страница 63: ...Falcon GC Appendix Teledyne Analytical Instruments 53 ...
Страница 64: ...Appendix Falcon GC 54 Teledyne Analytical Instruments ...
Страница 65: ...Falcon GC Appendix Teledyne Analytical Instruments 55 ...
Страница 66: ...Appendix Falcon GC 56 Teledyne Analytical Instruments ...
Страница 67: ...Falcon GC Appendix Teledyne Analytical Instruments 57 ...