Falcon GC
Operation
Teledyne Analytical Instruments
17
Operation
4.1 Starting the GC
4.1.1 Turning on the GC Modules
Now that the method file has been downloaded to the GC you are
ready to activate the Falcon modules and heat the instrument.
From the home screen on the Falcon touch screen, simply touch each
module icon individually and turn on their temperature and pressure
controllers. Press “Back” to return to the home screen from each
module. The order is not important as long as you turn on the injector
oven module and carrier gas first. Conversely, you can also access each
module by pressing Menu→Modules.
From the home screen you should be able to monitor the current
temperature or pressure of each module. You will know the setpoint has
been reached when the indicator next to the specific temperature or
pressure turns green. Yellow is below the setpoint and red is over
setpoint.
Содержание FALCON GC
Страница 10: ...Falcon GC x Teledyne Analytical Instruments Blank Page ...
Страница 15: ...Falcon GC Preparatio n Teledyne Analytical Instruments 5 3 Click on Report ...
Страница 16: ...Preparation Falcon GC 6 Teledyne Analytical Instruments Blank Page ...
Страница 42: ...Maintenance Troubleshooting Falcon GC 32 Teledyne Analytical Instruments ...
Страница 58: ...Maintenance Troubleshooting Falcon GC 48 Teledyne Analytical Instruments Figure 5 1 Glow Plug Wire ...
Страница 60: ...Maintenance Troubleshooting Falcon GC 50 Teledyne Analytical Instruments Figure 5 2 FID Diagram ...
Страница 62: ...Appendix Falcon GC 52 Teledyne Analytical Instruments A 3 Drawings ...
Страница 63: ...Falcon GC Appendix Teledyne Analytical Instruments 53 ...
Страница 64: ...Appendix Falcon GC 54 Teledyne Analytical Instruments ...
Страница 65: ...Falcon GC Appendix Teledyne Analytical Instruments 55 ...
Страница 66: ...Appendix Falcon GC 56 Teledyne Analytical Instruments ...
Страница 67: ...Falcon GC Appendix Teledyne Analytical Instruments 57 ...