Oxygen Analyzer
Operation
Teledyne Analytical Instruments
55
with a set of form “C" contacts rated for 3 amperes resistive load at 250
VAC. See Figure 3-5 in Chapter 3,
Installation
and/or the
Interconnection Diagram included at the back of this manual for relay
terminal connections.
The system failure alarm has a fixed configuration as described in
Chapter 3
Installation
.
The concentration alarms can be configured from the front panel as
either
high
or
low
alarms by the operator. The alarm modes can be set as
latching
or
non-latching
, and either
failsafe
or
non-failsafe
, or, they can
be
defeated
altogether. The setpoints for the alarms are also established
using this function.
Decide how your alarms should be configured. The choice will
depend upon your process. Consider the following four points:
1. Which if any of the alarms are to be high alarms and which if
any are to be low alarms?
Setting an alarm as HIGH triggers the alarm when the oxygen
concentration rises above the setpoint. Setting an alarm as LOW
triggers the alarm when the oxygen concentration falls below
the setpoint.
Decide whether you want the alarms to be set as:
•
Both high (high and high-high) alarms, or
•
One high and one low alarm, or
•
Both low (low and low-low) alarms.
2. Are either or both of the alarms to be configured as failsafe?
In failsafe mode, the alarm relay de-energizes in an alarm
condition. For non-failsafe operation, the relay is energized in
an alarm condition. You can set either or both of the
concentration alarms to operate in failsafe or non-failsafe mode.
3. Are either of the alarms to be latching?
In latching mode, once the alarm or alarms trigger, they will
remain in the alarm mode even if process conditions revert back
to non-alarm conditions. This mode requires an alarm to be
recognized before it can be reset. In the non-latching mode, the
alarm status will terminate when process conditions revert to
non-alarm conditions.
Содержание BDS 3000
Страница 14: ......
Страница 30: ...Operational Theory BDS 3000 Teledyne Analytical Instruments 16 Figure 2 7 Electronic Component Location...
Страница 32: ...Operational Theory BDS 3000 Teledyne Analytical Instruments 18...
Страница 46: ...Installation BDS 3000 Teledyne Analytical Instruments 32...
Страница 49: ...Oxygen Analyzer Operation Teledyne Analytical Instruments 35 Figure 4 1 Hierarchy of Functions and Subfunctions...
Страница 76: ...Operation BDS 3000 Teledyne Analytical Instruments 62 Figure 4 2 Analyzer Power up Sequence...
Страница 77: ...Oxygen Analyzer Operation Teledyne Analytical Instruments 63 Figure 4 3 Analyzer Span Sequence...
Страница 78: ...Operation BDS 3000 Teledyne Analytical Instruments 64 Figure 4 4 Analyzer Zero Sequence...
Страница 86: ...Maintenance BDS 3000 Teledyne Analytical Instruments 72...
Страница 92: ...Appendix BDS 3000 Teledyne Analytical Instruments 78...
Страница 96: ...Appendix MSDS Teledyne Analytical Instruments 82...