Operational Theory
BDS 3000
Teledyne Analytical Instruments
10
the sensing electrode. The reference electrode provides a potential
reference for both the sensing electrode and the 2nd working electrode.
NOTE: BDS technology and sensor is a patent pending
technology of Teledyne Analytical Instruments in the
United State of America as well as many foreign countries.
To learn more about BDS technology, please visit TAI’s web page at
http://www.Teledyne-AI.com
To learn more about potentiostat, visit Electochemical Society’s web
page at http://www.electrochem.org
2.2.2 Gas Flow Rate
The output from a BDS oxygen sensor is relatively insensitive to
change of gas flow rate if operated in the range of 1 - 3 SCFH (in
nitrogen). The output drops when the flow rate is below 1 SCFH. Figure
2.2 is a typical curve showing the sensor outputs at different flow rate.
Figure 2.2: BDS sensor output at different gas flow rate
Содержание BDS 3000
Страница 14: ......
Страница 30: ...Operational Theory BDS 3000 Teledyne Analytical Instruments 16 Figure 2 7 Electronic Component Location...
Страница 32: ...Operational Theory BDS 3000 Teledyne Analytical Instruments 18...
Страница 46: ...Installation BDS 3000 Teledyne Analytical Instruments 32...
Страница 49: ...Oxygen Analyzer Operation Teledyne Analytical Instruments 35 Figure 4 1 Hierarchy of Functions and Subfunctions...
Страница 76: ...Operation BDS 3000 Teledyne Analytical Instruments 62 Figure 4 2 Analyzer Power up Sequence...
Страница 77: ...Oxygen Analyzer Operation Teledyne Analytical Instruments 63 Figure 4 3 Analyzer Span Sequence...
Страница 78: ...Operation BDS 3000 Teledyne Analytical Instruments 64 Figure 4 4 Analyzer Zero Sequence...
Страница 86: ...Maintenance BDS 3000 Teledyne Analytical Instruments 72...
Страница 92: ...Appendix BDS 3000 Teledyne Analytical Instruments 78...
Страница 96: ...Appendix MSDS Teledyne Analytical Instruments 82...