Automation
Section 3 The AutoSampler Module
3-11
3.1.2 Establishing
Connections
To establish the necessary connections, follow the steps below:
1. Place the AutoSampler Module on the right side of the EZ
Prep.
2. Remove the Luer fitting from the EZ Prep inject valve.
Figure 3-18 Removing the default EZ Prep injector port
3. Locate the supplied sample probe tubing and fitting pack-
age (
P
/
N
60-5234-651).
4. Unscrew the attached plastic ferrule in this fitting package
until the plastic ferrule is almost completely detached from
the steel fitting.
5. Attach the steel fitting in this package to the EZ Prep
inject valve waste port. This steel fitting should be hand
tight.
Figure 3-19 Sample probe tube fitting connected to the EZ
Prep inject valve
6. Turn the plastic ferrule until it is attached hand tight to
the valve.
7. Turn the steel fitting
1
/
2
turn tighter.