ADFM® Flow Meter
Section 2 Setup and Installation
2-57
Table 2-2 BIT Hexadecimal Code Definitions
Code
Definition
Explanation
01nnnnnn
Autostart:
Unit has been restarted or restarted itself.
80nnnnnn
Modem dial failure
40nnnnnn
Modem listen failure
20nnnnnn
Recorder failure
10nnnnnn
Recorder full
08nnnnnn
Transducer communication failure
04nnnnnn
Power time fail:
Motherboard failure
02nnnnnn
Fifo fail:
008nnnnn
DeMod LCA fail:
00000040
Intrup Err:
00000020
00000010
00000008
Divo Err:
00000004
Inst Err:
00000002
Addr Err:
00000001
Bus Err:
Possible motherboard failure, but may
correct itself upon restart.
004nnnnn
Comm Timeout: Communication times out in the ADFM after several minutes of no activity.
To reestablish communication, disconnect the ADFM from the computer and then reconnect.
002nnnnn
Pressure fail: Failure of secondary depth sensor if it is a pressure type sensor;
no code for other sensor types.
00000080
Cold WU (wakeup): Manual restart occurred while ADFM was connected to computer.
Note
Each code indicates one error or event type. However, if two or more events are present, the code
will appear as the sum of the most significant digits of their corresponding codes.
Example: If the ADFM has been restarted and also experienced a transducer communication failure,
the code displayed would be:
09nnnnnn
.
Содержание ADFM Hot Tap
Страница 2: ......
Страница 4: ......
Страница 20: ...ADFM Flow Meter Section 1 Introduction 1 10 ...
Страница 85: ...ADFM Flow Meter Section 2 Setup and Installation 2 65 Figure 2 73 Viewing data in Expert View ...
Страница 86: ...ADFM Flow Meter Section 2 Setup and Installation 2 66 ...
Страница 98: ...ADFM Flow Meter Section 3 Maintenance and Troubleshooting 3 12 ...
Страница 114: ......
Страница 116: ......