Teledyne ACCQPrep AI 5 Скачать руководство пользователя страница 1

Installing the ACCQ

Prep

AI 5

AutoInjector

Overview

This instruction sheet will guide you through the

installation of the ACCQ

Prep

AI 5 AutoInjector into the

ACCQ

Prep

HP125 system.

Tools Required:

1/4” wrench

#2 Phillips screwdriver

#1 Phillips screwdriver

Side cutters

Parts List:

Installation

1. Turn O

FF

power to the ACCQ

Prep

HP125 system.

2. Disconnect the power cord from the back of the

unit.

3. Remove the cover panel on the right side of the

instrument (Figure 1, L).

4. Carefully cut the plastic tie securing the AutoIn-

jector power supply cable to the back of the
panel (Figure 1, R).

Figure 1: Cover panel (L) and cable tie (R)

5. Connect the AutoInjector power cable to the

port on the back of the syringe drive and secure
with the provided panhead screws and internal
lock washers (Figure 2).

Note

If the power supply cable isn’t long enough, you may
need to access the back of the AutoInjector through
the back panel of ACCQ

Prep

HP125 system.

Figure 2: Location of screw placement on the
syringe drive

6. Once the power supply cable is secured, you

may

now slide the AutoInjector Module into

the ACCQ

Prep

system. (Figure 3).

Figure 3: AutoInjector partially installed

7. Using the screws (removed in Step 3) attach the

AutoInjector assembly into place.

8. The back panel may be reinstalled on the system

if it was necessary to remove it.

Part

Part Number

AutoInjector

60-5237-098

Panhead Screws

231514303

Lock Washers

233210400

72” Black Tubing

023050309

15” Teflon Tubing

023050204

Cable tie

489010100

Instruction Sheet #69-5233-836

March 2018