
4250 Flow Meter
Section 2 Programming
2-49
2.4.7 Step 4 - Reset Totalizer
If you select
step 4
, the following will appear:
This step allows you to reset the flow meter’s internal flow total-
izers. All models have the capability of maintaining a separate
totalizer for the time the sampler is enabled through the
sampler enabling
feature (
step 6
). The 4250 flow meter,
because it can measure forward and reverse flows, maintains
three separate totalizers
: a separate totalizer for
each
direction
and a totalizer for the
difference
between forward
and reverse flows.
If you select NO, the flow meter will advance to the next step. If
you select YES, the flow meter will reset the internal totalizer to
zero.
Then, if you have sampler enabling turned on (
step 6
), the flow
meter will ask you whether you want to reset the totalizer for the
total flow during the time the sampler was enabled. (See
step 6
for an explanation of sampler enabling.)
Then:
2.4.8 Step 5 - Sampler
Pacing
If you select
step 5
, the following will appear:
This step determines how the flow meter will signal an asso-
ciated automatic wastewater sampler to take a sample. The flow
meter and sampler must be connected together with a cable. The
flow meter sends flow pulses to the sampler. The sampler uses
these flow pulses as counts. When an appropriate number of flow
pulses has been received by the sampler, it will take a sample.
VOLUME will not appear if the flow meter is measuring level
only, or is only being used for parameter monitoring.
The Flowlink menu option will not appear unless Flowlink
software is installed and pacing has been downloaded from
Flowlink.
If you select DISABLE for sampler pacing, the flow meter will be
effectively disconnected from the sampler. The sampler will run
its program as if there were no flow meter. Selection of DISABLE
RESET TOTALIZER
• YES • • NO •
FLOW TOTALIZER:
XX
CF
FORWARD
XX
REVERSE
X
CF
ENABLE TOTALIZER
XX
CF
(or other units of measure)
PRESS 'ENTER
RESET SAMPLER ENABLE TOTALIZER
• YES • • NO •
SAMPLER PACING
• DISABLE • • (VOLUME) • • (FLOWLINK) • • CONDITIONAL •
Содержание 4250
Страница 2: ......
Страница 4: ......
Страница 12: ...4250 Flow Meter Table of Contents x ...
Страница 28: ...4250 Flow Meter Section 1 Introduction 1 16 ...
Страница 90: ...4250 Flow Meter Section 2 Programming 2 62 ...
Страница 140: ...4250 Flow Meter Section 4 Options and Accessories 4 30 ...
Страница 160: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 2 ...
Страница 161: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 3 ...
Страница 166: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 8 ...
Страница 184: ...4250 Flow Meter Appendix C General Safety Procedures C 10 ...
Страница 191: ...4250 Flow Meter Appendix D Material Safety Data Sheets D 7 ...
Страница 192: ...4250 Flow Meter Appendix D Material Safety Data Sheets D 8 ...
Страница 196: ......
Страница 198: ......
Страница 200: ......
Страница 202: ...Warranty ...
Страница 204: ......